OutageID |
SiteID |
Failure Type |
Start Date |
End Date |
Description |
Resolution Description |
916 |
str |
Power Failure |
2024-06-09 22:58:00 |
|
UPS is non-responsive |
|
884 |
str |
Hardware - Instrument |
2023-10-18 02:15:00 |
2024-01-26 00:00:00 |
Receiver modulator box malfunction |
Cold solder joint repair |
907 |
str |
Hardware - Instrument |
2023-10-18 02:15:00 |
2024-03-13 21:15:00 |
Coupler disabled |
Coupler enabled |
857 |
str |
Hardware - Instrument |
2023-05-13 02:00:00 |
2023-07-12 17:54:00 |
Low battery voltage preventing overnight datalogger operations |
Battery replaced |
855 |
str |
Hardware - Instrument |
2023-04-23 22:16:00 |
2023-07-12 17:54:00 |
Temperature/relative-humidity sensor malfunction |
Sensor replaced |
856 |
str |
Hardware - Instrument |
2023-04-14 07:34:00 |
2023-07-12 17:54:00 |
Pressure sensor malfunction |
Sensor replaced |
612 |
str |
Power Failure |
2018-11-21 16:00:00 |
2019-04-24 21:40:00 |
Power issue confirmed. |
Power restored. |
590 |
str |
Hardware - Instrument |
2018-04-30 21:12:00 |
2018-05-10 21:00:00 |
Can't connect to datalogger. Data cable from PC to datalogger was damaged during groundskeeping work. |
Cable replaced. |
558 |
str |
Communication - Cellular |
2017-10-09 13:00:00 |
2017-10-11 03:45:00 |
Router unreachable |
Cellular service restored. Wildfire in the vicinity. |
741 |
str |
Hardware - Instrument |
2016-11-29 23:00:00 |
2022-06-14 02:40:00 |
100-cm soil reflectometer data is suspect |
sensor verified to be working correctly; suspect data may be related to abnormal soil and/or water behaviors |
503 |
str |
Communication - Cellular |
2016-10-06 12:00:00 |
2016-10-10 00:00:00 |
Lost connectivity to radar PC via cell modem. |
Comms restored. |
486 |
str |
Communication - Cellular |
2016-05-28 14:00:00 |
2016-06-01 22:45:00 |
Unable to reach router. Possible power outage or router issue. |
UPS reset and power restored. |
414 |
str |
Communication - Cellular |
2014-10-24 00:00:00 |
2014-11-05 21:00:00 |
Router not responding. Could be a router and/or power issue. |
Contact rebooted system. |
369 |
str |
Hardware - Instrument |
2014-03-07 00:00:00 |
2014-04-07 19:00:00 |
Datalogger is dead. Needs to be replaced. |
Datalogger replaced. |
363 |
str |
Hardware - Instrument |
2014-02-07 00:00:00 |
2014-02-27 00:00:00 |
Tipping bucket gauge most likely clogged as seen from linear rain trace in plot. |
clogged fixed |
188 |
str |
Hardware - Instrument |
2012-02-08 00:00:00 |
2012-02-15 00:00:00 |
Faulty Temperature/RH sensor. Reporting values too low. Sending contact replacement sensor. |
Replaced Temperature/RH sensor. |